Fully Integrated Flowmeter |
Emerson Rosemount 3051SFA with Measurement Types 1-4 | Emerson Rosemount 3051SFA Measurement Type D |
Accuracy | •Mass flow up to 0.8% of mass flow rate, 14:1 turndown •Classic MV up to ±0.85 % of flow rate, 8:1 flow turndown | •Volumetric flow up to 0.8% of volumetric flow rate, 14:1 turndown •Classic MV up to ±0.85% of volumetric flow rate, 8:1 turndown |
Repeatability | ± 0.1 % | ± 0.1 % |
Output | •4-20mA signal based on HART® Communication •WirelessHART with Smart Wireless THUM adapter | •4-20mA signal based on HART® Communication •WirelessHART •FOUNDATION™ fieldbus |
Long Term Stability | Up to 15-year stability | Up to 15-year stability |
Warranty | Up to 15-year limited warranty | Up to 15-year limited warranty |
Diagnostics | •User-writable process temperature sensor limits •Sensor Temperature as an output •Instant Flow Verification | Advanced HART Diagnostic Suite |
Line Size | 2-in. to 8-in. (50 to 200 mm) – Sensor Size 1 6-in. to 36-in. (150 to 900 mm) – Sensor Size 2 12-in. to 72-in. (300 to 1800 mm) – Sensor Size 3 | 2-in. to 8-in. (50 to 200 mm) – Sensor Size 1 6-in. to 36-in. (150 to 900 mm) – Sensor Size 2 12-in. to 72-in. (300 to 1800 mm) – Sensor Size 3 |